초록
The increasing global prevalence of diabetes necessitates the development of accurate, reliable, and accessible glucose monitoring devices, particularly for point-of-care testing. In this study, we developed an electrochemical glucose biosensor using electrospun polyvinyl alcohol (PVA) nanofibers. Among three immobilization strategies, dip-coating glucose oxidase (GOx) exhibits superior electrochemical reversibility and a lower operating potential compared to drop-casting and electrospinning techniques. By integrating reduced graphene oxide (rGO) to enhance the conductivity of the polymer matrix, the resulting dip-GOx/rGO/PVA biosensor demonstrated significantly enhanced electron transfer kinetics () and an increase in sensitivity from 35.41 & micro;A mM-1 cm-2 to 51.36 & micro;A mM-1 cm-2. The sensor maintained a linear response range of 3-9 mM with a limit of detection of 1.23 mM, covering the critical physiological glucose levels. Although less sensitive than noble-metal sensors, this platform is cost-effective, eco-friendly, and highly biocompatible, making it a sustainable candidate for disposable biosensors.
키워드
- 제목
- Electrospun rGO/PVA Nanofibrous Biosensors: Impact of Enzyme Immobilization on Electrochemical Performance
- 저자
- Vo, Hoai Phuong; Nguyen, Minh Phong; Nguyen, Minh Doan; Nguyen, Thi Dao; Nguyen, Hai Dang; Pham, Ngoc Thao; Nguyen, Duc San; Can, Thi Thu Thuy; Nguyen, Van Quang; Pham, Duy Tho; Phan, The-Long; Nguyen, Tuan Canh
- 발행일
- 2026-07-21
- 유형
- Article; Early Access
- 언어
- ENG
- 출판사
- SPRINGER
- 발행국가
- 미국
- ISSN
- E 1543-186X
P 0361-5235