Dual-site PGM-based point-of-care detection of EGFR exon 19 deletion with an endogenous internal control
  • Jeon, Hyo Won
  • Han, Hyogu
  • Yang, Jieun
  • Ahn, Jun Ki
Citations

WEB OF SCIENCE

0
Citations

SCOPUS

0

초록

Rapid and reliable detection of EGFR exon 19 deletion (19del) is important for precision treatment of non-small cell lung cancer, yet current genotyping methods remain poorly suited for point-of-care testing (POCT). This limitation is particularly significant in signal-off assays, where reaction failure can be misinterpreted as a true negative result. Here, we present a dual-site personal glucose meter (PGM)-readout platform for minimal-instrumentation analysis of EGFR 19del. In this system, target-dependent flap endonuclease 1 (FEN1) cleavage generates adenosine monophosphate (AMP), which triggers a glucose-consuming kinase cascade and enables direct quantitative readout using a commercial PGM without optical instrumentation or nucleic acid amplification. To improve interpretability, the platform incorporates an endogenous internal control through a dual-site design consisting of a mutation-discriminating detection site and a conserved control site, thereby enabling reliable interpretation of signal-off readouts while confirming sample adequacy. The platform achieved sensitive detection with a limit of detection of approximately 27 pM and discriminated mutation fractions down to ∼0.44% in wild-type/mutant mixtures. Its feasibility was further demonstrated using cell-derived genomic DNA and serum-containing samples. These results indicate that the proposed platform provides an accessible and practical strategy for decentralized EGFR 19del testing.

키워드

Dual-site sensor platformEpidermal growth factor receptor (EGFR)Flap endonuclease 1Kinase cascadePersonal glucose meterMUTATIONSCANCERASSAY
제목
Dual-site PGM-based point-of-care detection of EGFR exon 19 deletion with an endogenous internal control
저자
Jeon, Hyo WonHan, HyoguYang, JieunAhn, Jun Ki
DOI
10.1016/j.bios.2026.118674
발행일
2026-08-01
유형
Article
저널명
Biosensors and Bioelectronics
305