Hierarchical deep learning pipeline for robust cervical parameter measurement in radiographs with C7 obscuration

Citations

WEB OF SCIENCE

0
Citations

SCOPUS

0

초록

We developed and externally validated a hierarchical deep learning pipeline that automates cervical sagittal measurements, explicitly addressing C7 obscuration on lateral radiographs. The model combines a global keypoint detector with C2/C7 specialists localized via a multilayer perceptron to refine landmarks on high-resolution patches. Trained on 5604 images and tested internally and on a challenging external cohort enriched for C7 obscuration (82%), it achieved excellent agreement with ground truth. Externally, intraclass correlation coefficients (ICCs) were 0.97 for lordosis (mean absolute error [MAE] 2.6 degrees), >0.99 for C2 slope (MAE 0.8 degrees), and 0.93 for C7 slope (MAE 2.3 degrees), with minimal bias and narrower limits of agreement than a single-stage baseline. The model showed near-perfect repeatability (ICC > 0.99) and higher artificial intelligence-expert agreement (ICC 0.81-0.84) for C7 slope than inter-expert reliability (ICC 0.67). In failure cases, the pipeline corrected large global model errors (e.g., 10.22 degrees- 0.22 degrees). This robust, coarse-to-fine approach advances reliable, generalizable cervical alignment assessment in real-world conditions.

키워드

ALIGNMENT
제목
Hierarchical deep learning pipeline for robust cervical parameter measurement in radiographs with C7 obscuration
저자
Kang, Dong-HoPark, Se-JunPark, Jin-SungPark, HyeonsuLee, Chong-Suh
DOI
10.1038/s41746-026-02455-2
발행일
2026-02-19
유형
Article
저널명
NPJ DIGITAL MEDICINE
9
1